Quantitative Radioisotope Measurements with the NSF-Arizona Regional Accelerator Facility
Results of tests on the tandem accelerator mass spectrometer (TAMS) at the University of Arizona are presented. These results demonstrate: (a) measurements of 14C/13C ratios with precisions of a few percent can be made in a period of one to several hours; (b) measurements with precisions of 0.5% hav...
Gespeichert in:
Veröffentlicht in: | IEEE Trans. Nucl. Sci.; (United States) 1983-01, Vol.30 (2), p.1371-1373 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Results of tests on the tandem accelerator mass spectrometer (TAMS) at the University of Arizona are presented. These results demonstrate: (a) measurements of 14C/13C ratios with precisions of a few percent can be made in a period of one to several hours; (b) measurements with precisions of 0.5% have been made in which the uncertainties were mainly statistical and in which contributions to the uncertainty of machine fluctuations were negligible; (c) precise measurements of the ratio of 14C/13C in samples of N.B.S. oxalic acid and of 1890 wood are consistent with the accepted value of that ratio; (d) the real signal from a 44,000 year old sample is equal to the background rate produced from a dead carbon sample. In addition, results of some measurements on archaeological samples are presented. |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/TNS.1983.4332535 |