PACEMAKER FAILURE ON INDUCTION OF ANAESTHESIA

A patient with a permanent pacemaker presented for repair of a strangulated hernia. During induction of anaesthesia, the pacemaker generator stopped discharging, thus causing cardiac arrest. The likely cause of the generator failure was inhibition by suxamethonium- induced muscle fasciculations. Fol...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1991-04, Vol.66 (4), p.509-512
1. Verfasser: FINFER, S.R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A patient with a permanent pacemaker presented for repair of a strangulated hernia. During induction of anaesthesia, the pacemaker generator stopped discharging, thus causing cardiac arrest. The likely cause of the generator failure was inhibition by suxamethonium- induced muscle fasciculations. Following defibrillation, an increase in stimulation threshold necessitated urgent insertion of a transvenous pacing system. It is suggested that, when suxamethonium is to be used in a patient with a permanent pacemaker, consideration should be given to reprogramming the pacemaker to asynchronous mode before induction of anaesthesia. If a patient with a pacemaker requires defibrillation, an acute increase in stimulation threshold may result and cause loss of capture. Rapid insertion of a transvenous pacing system may be necessary.
ISSN:0007-0912
1471-6771
DOI:10.1093/bja/66.4.509