Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study

The purpose of this study is to evaluate the impact of a novel tissue characterization method using dual-energy over single-energy computed tomography (DECT and SECT) on Monte Carlo (MC) dose calculations for low-dose rate (LDR) prostate brachytherapy performed in a patient like geometry. A virtual...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2018-01, Vol.63 (2), p.025013-025013
Hauptverfasser: Remy, Charlotte, Lalonde, Arthur, Béliveau-Nadeau, Dominic, Carrier, Jean-François, Bouchard, Hugo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 025013
container_issue 2
container_start_page 025013
container_title Physics in medicine & biology
container_volume 63
creator Remy, Charlotte
Lalonde, Arthur
Béliveau-Nadeau, Dominic
Carrier, Jean-François
Bouchard, Hugo
description The purpose of this study is to evaluate the impact of a novel tissue characterization method using dual-energy over single-energy computed tomography (DECT and SECT) on Monte Carlo (MC) dose calculations for low-dose rate (LDR) prostate brachytherapy performed in a patient like geometry. A virtual patient geometry is created using contours from a real patient pelvis CT scan, where known elemental compositions and varying densities are overwritten in each voxel. A second phantom is made with additional calcifications. Both phantoms are the ground truth with which all results are compared. Simulated CT images are generated from them using attenuation coefficients taken from the XCOM database with a 100 kVp spectrum for SECT and 80 and 140Sn kVp for DECT. Tissue segmentation for Monte Carlo dose calculation is made using a stoichiometric calibration method for the simulated SECT images. For the DECT images, Bayesian eigentissue decomposition is used. A LDR prostate brachytherapy plan is defined with 125I sources and then calculated using the EGSnrc user-code Brachydose for each case. Dose distributions and dose-volume histograms (DVH) are compared to ground truth to assess the accuracy of tissue segmentation. For noiseless images, DECT-based tissue segmentation outperforms the SECT procedure with a root mean square error (RMS) on relative errors on dose distributions respectively of 2.39% versus 7.77%, and provides DVHs closest to the reference DVHs for all tissues. For a medium level of CT noise, Bayesian eigentissue decomposition still performs better on the overall dose calculation as the RMS error is found to be of 7.83% compared to 9.15% for SECT. Both methods give a similar DVH for the prostate while the DECT segmentation remains more accurate for organs at risk and in presence of calcifications, with less than 5% of RMS errors within the calcifications versus up to 154% for SECT. In a patient-like geometry, DECT-based tissue segmentation provides dose distributions with the highest accuracy and the least bias compared to SECT. When imaging noise is considered, benefits of DECT are noticeable if important calcifications are found within the prostate.
doi_str_mv 10.1088/1361-6560/aaa30c
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1088_1361_6560_aaa30c</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>1979171654</sourcerecordid><originalsourceid>FETCH-LOGICAL-c336t-5c4b851583308b17f147dbb2577a30aeded830aa9558598c9e4d2505d037329c3</originalsourceid><addsrcrecordid>eNp1kD1PwzAURS0EoqWwMyGPDATsOI4dNlQ-pSKWMluO81JSJXGwHaH8e1wV2Jie9Hzele9B6JySa0qkvKEsp0nOc3KjtWbEHKD53-oQzQlhNCko5zN04v2WEEplmh2jWVqkORGpmCN3b33TQXCNwU03aBOwrXE16jaBHtxmwss1Do33I2APmw76oENje1xbh1v79UsNzvr4Arh02nxM4QOcHqZbrPGr7eN6qV1rsQ9jNZ2io1q3Hs5-5gK9Pz6sl8_J6u3pZXm3SgxjeUi4yUrJKZeMEVlSUdNMVGWZciFiVQ0VVDJOXXAueSFNAVmVcsIrwgRLC8MW6HKfG__2OYIPqmu8gbbVPdjRK1qIggqa8yyiZI-aWMM7qNXgmk67SVGidqbVTqvaaVV70_Hk4id9LDuo_g5-1Ubgag80dlBbO7o-lv0_7xsIM4h5</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1979171654</pqid></control><display><type>article</type><title>Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study</title><source>MEDLINE</source><source>IOP Publishing Journals</source><source>Institute of Physics (IOP) Journals - HEAL-Link</source><creator>Remy, Charlotte ; Lalonde, Arthur ; Béliveau-Nadeau, Dominic ; Carrier, Jean-François ; Bouchard, Hugo</creator><creatorcontrib>Remy, Charlotte ; Lalonde, Arthur ; Béliveau-Nadeau, Dominic ; Carrier, Jean-François ; Bouchard, Hugo</creatorcontrib><description>The purpose of this study is to evaluate the impact of a novel tissue characterization method using dual-energy over single-energy computed tomography (DECT and SECT) on Monte Carlo (MC) dose calculations for low-dose rate (LDR) prostate brachytherapy performed in a patient like geometry. A virtual patient geometry is created using contours from a real patient pelvis CT scan, where known elemental compositions and varying densities are overwritten in each voxel. A second phantom is made with additional calcifications. Both phantoms are the ground truth with which all results are compared. Simulated CT images are generated from them using attenuation coefficients taken from the XCOM database with a 100 kVp spectrum for SECT and 80 and 140Sn kVp for DECT. Tissue segmentation for Monte Carlo dose calculation is made using a stoichiometric calibration method for the simulated SECT images. For the DECT images, Bayesian eigentissue decomposition is used. A LDR prostate brachytherapy plan is defined with 125I sources and then calculated using the EGSnrc user-code Brachydose for each case. Dose distributions and dose-volume histograms (DVH) are compared to ground truth to assess the accuracy of tissue segmentation. For noiseless images, DECT-based tissue segmentation outperforms the SECT procedure with a root mean square error (RMS) on relative errors on dose distributions respectively of 2.39% versus 7.77%, and provides DVHs closest to the reference DVHs for all tissues. For a medium level of CT noise, Bayesian eigentissue decomposition still performs better on the overall dose calculation as the RMS error is found to be of 7.83% compared to 9.15% for SECT. Both methods give a similar DVH for the prostate while the DECT segmentation remains more accurate for organs at risk and in presence of calcifications, with less than 5% of RMS errors within the calcifications versus up to 154% for SECT. In a patient-like geometry, DECT-based tissue segmentation provides dose distributions with the highest accuracy and the least bias compared to SECT. When imaging noise is considered, benefits of DECT are noticeable if important calcifications are found within the prostate.</description><identifier>ISSN: 0031-9155</identifier><identifier>ISSN: 1361-6560</identifier><identifier>EISSN: 1361-6560</identifier><identifier>DOI: 10.1088/1361-6560/aaa30c</identifier><identifier>PMID: 29260727</identifier><identifier>CODEN: PHMBA7</identifier><language>eng</language><publisher>England: IOP Publishing</publisher><subject>Brachytherapy ; Brachytherapy - methods ; dual-energy CT ; Humans ; Male ; Monte Carlo dose calculation ; Monte Carlo Method ; Phantoms, Imaging ; Prostatic Neoplasms - diagnostic imaging ; Prostatic Neoplasms - radiotherapy ; Radiotherapy Dosage ; Radiotherapy Planning, Computer-Assisted - methods ; tissue characterization ; Tomography, X-Ray Computed - methods</subject><ispartof>Physics in medicine &amp; biology, 2018-01, Vol.63 (2), p.025013-025013</ispartof><rights>2018 Institute of Physics and Engineering in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c336t-5c4b851583308b17f147dbb2577a30aeded830aa9558598c9e4d2505d037329c3</citedby><cites>FETCH-LOGICAL-c336t-5c4b851583308b17f147dbb2577a30aeded830aa9558598c9e4d2505d037329c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://iopscience.iop.org/article/10.1088/1361-6560/aaa30c/pdf$$EPDF$$P50$$Giop$$H</linktopdf><link.rule.ids>314,776,780,27901,27902,53821,53868</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/29260727$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Remy, Charlotte</creatorcontrib><creatorcontrib>Lalonde, Arthur</creatorcontrib><creatorcontrib>Béliveau-Nadeau, Dominic</creatorcontrib><creatorcontrib>Carrier, Jean-François</creatorcontrib><creatorcontrib>Bouchard, Hugo</creatorcontrib><title>Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study</title><title>Physics in medicine &amp; biology</title><addtitle>PMB</addtitle><addtitle>Phys. Med. Biol</addtitle><description>The purpose of this study is to evaluate the impact of a novel tissue characterization method using dual-energy over single-energy computed tomography (DECT and SECT) on Monte Carlo (MC) dose calculations for low-dose rate (LDR) prostate brachytherapy performed in a patient like geometry. A virtual patient geometry is created using contours from a real patient pelvis CT scan, where known elemental compositions and varying densities are overwritten in each voxel. A second phantom is made with additional calcifications. Both phantoms are the ground truth with which all results are compared. Simulated CT images are generated from them using attenuation coefficients taken from the XCOM database with a 100 kVp spectrum for SECT and 80 and 140Sn kVp for DECT. Tissue segmentation for Monte Carlo dose calculation is made using a stoichiometric calibration method for the simulated SECT images. For the DECT images, Bayesian eigentissue decomposition is used. A LDR prostate brachytherapy plan is defined with 125I sources and then calculated using the EGSnrc user-code Brachydose for each case. Dose distributions and dose-volume histograms (DVH) are compared to ground truth to assess the accuracy of tissue segmentation. For noiseless images, DECT-based tissue segmentation outperforms the SECT procedure with a root mean square error (RMS) on relative errors on dose distributions respectively of 2.39% versus 7.77%, and provides DVHs closest to the reference DVHs for all tissues. For a medium level of CT noise, Bayesian eigentissue decomposition still performs better on the overall dose calculation as the RMS error is found to be of 7.83% compared to 9.15% for SECT. Both methods give a similar DVH for the prostate while the DECT segmentation remains more accurate for organs at risk and in presence of calcifications, with less than 5% of RMS errors within the calcifications versus up to 154% for SECT. In a patient-like geometry, DECT-based tissue segmentation provides dose distributions with the highest accuracy and the least bias compared to SECT. When imaging noise is considered, benefits of DECT are noticeable if important calcifications are found within the prostate.</description><subject>Brachytherapy</subject><subject>Brachytherapy - methods</subject><subject>dual-energy CT</subject><subject>Humans</subject><subject>Male</subject><subject>Monte Carlo dose calculation</subject><subject>Monte Carlo Method</subject><subject>Phantoms, Imaging</subject><subject>Prostatic Neoplasms - diagnostic imaging</subject><subject>Prostatic Neoplasms - radiotherapy</subject><subject>Radiotherapy Dosage</subject><subject>Radiotherapy Planning, Computer-Assisted - methods</subject><subject>tissue characterization</subject><subject>Tomography, X-Ray Computed - methods</subject><issn>0031-9155</issn><issn>1361-6560</issn><issn>1361-6560</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kD1PwzAURS0EoqWwMyGPDATsOI4dNlQ-pSKWMluO81JSJXGwHaH8e1wV2Jie9Hzele9B6JySa0qkvKEsp0nOc3KjtWbEHKD53-oQzQlhNCko5zN04v2WEEplmh2jWVqkORGpmCN3b33TQXCNwU03aBOwrXE16jaBHtxmwss1Do33I2APmw76oENje1xbh1v79UsNzvr4Arh02nxM4QOcHqZbrPGr7eN6qV1rsQ9jNZ2io1q3Hs5-5gK9Pz6sl8_J6u3pZXm3SgxjeUi4yUrJKZeMEVlSUdNMVGWZciFiVQ0VVDJOXXAueSFNAVmVcsIrwgRLC8MW6HKfG__2OYIPqmu8gbbVPdjRK1qIggqa8yyiZI-aWMM7qNXgmk67SVGidqbVTqvaaVV70_Hk4id9LDuo_g5-1Ubgag80dlBbO7o-lv0_7xsIM4h5</recordid><startdate>20180111</startdate><enddate>20180111</enddate><creator>Remy, Charlotte</creator><creator>Lalonde, Arthur</creator><creator>Béliveau-Nadeau, Dominic</creator><creator>Carrier, Jean-François</creator><creator>Bouchard, Hugo</creator><general>IOP Publishing</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20180111</creationdate><title>Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study</title><author>Remy, Charlotte ; Lalonde, Arthur ; Béliveau-Nadeau, Dominic ; Carrier, Jean-François ; Bouchard, Hugo</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c336t-5c4b851583308b17f147dbb2577a30aeded830aa9558598c9e4d2505d037329c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Brachytherapy</topic><topic>Brachytherapy - methods</topic><topic>dual-energy CT</topic><topic>Humans</topic><topic>Male</topic><topic>Monte Carlo dose calculation</topic><topic>Monte Carlo Method</topic><topic>Phantoms, Imaging</topic><topic>Prostatic Neoplasms - diagnostic imaging</topic><topic>Prostatic Neoplasms - radiotherapy</topic><topic>Radiotherapy Dosage</topic><topic>Radiotherapy Planning, Computer-Assisted - methods</topic><topic>tissue characterization</topic><topic>Tomography, X-Ray Computed - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Remy, Charlotte</creatorcontrib><creatorcontrib>Lalonde, Arthur</creatorcontrib><creatorcontrib>Béliveau-Nadeau, Dominic</creatorcontrib><creatorcontrib>Carrier, Jean-François</creatorcontrib><creatorcontrib>Bouchard, Hugo</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Physics in medicine &amp; biology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Remy, Charlotte</au><au>Lalonde, Arthur</au><au>Béliveau-Nadeau, Dominic</au><au>Carrier, Jean-François</au><au>Bouchard, Hugo</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study</atitle><jtitle>Physics in medicine &amp; biology</jtitle><stitle>PMB</stitle><addtitle>Phys. Med. Biol</addtitle><date>2018-01-11</date><risdate>2018</risdate><volume>63</volume><issue>2</issue><spage>025013</spage><epage>025013</epage><pages>025013-025013</pages><issn>0031-9155</issn><issn>1361-6560</issn><eissn>1361-6560</eissn><coden>PHMBA7</coden><abstract>The purpose of this study is to evaluate the impact of a novel tissue characterization method using dual-energy over single-energy computed tomography (DECT and SECT) on Monte Carlo (MC) dose calculations for low-dose rate (LDR) prostate brachytherapy performed in a patient like geometry. A virtual patient geometry is created using contours from a real patient pelvis CT scan, where known elemental compositions and varying densities are overwritten in each voxel. A second phantom is made with additional calcifications. Both phantoms are the ground truth with which all results are compared. Simulated CT images are generated from them using attenuation coefficients taken from the XCOM database with a 100 kVp spectrum for SECT and 80 and 140Sn kVp for DECT. Tissue segmentation for Monte Carlo dose calculation is made using a stoichiometric calibration method for the simulated SECT images. For the DECT images, Bayesian eigentissue decomposition is used. A LDR prostate brachytherapy plan is defined with 125I sources and then calculated using the EGSnrc user-code Brachydose for each case. Dose distributions and dose-volume histograms (DVH) are compared to ground truth to assess the accuracy of tissue segmentation. For noiseless images, DECT-based tissue segmentation outperforms the SECT procedure with a root mean square error (RMS) on relative errors on dose distributions respectively of 2.39% versus 7.77%, and provides DVHs closest to the reference DVHs for all tissues. For a medium level of CT noise, Bayesian eigentissue decomposition still performs better on the overall dose calculation as the RMS error is found to be of 7.83% compared to 9.15% for SECT. Both methods give a similar DVH for the prostate while the DECT segmentation remains more accurate for organs at risk and in presence of calcifications, with less than 5% of RMS errors within the calcifications versus up to 154% for SECT. In a patient-like geometry, DECT-based tissue segmentation provides dose distributions with the highest accuracy and the least bias compared to SECT. When imaging noise is considered, benefits of DECT are noticeable if important calcifications are found within the prostate.</abstract><cop>England</cop><pub>IOP Publishing</pub><pmid>29260727</pmid><doi>10.1088/1361-6560/aaa30c</doi><tpages>13</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0031-9155
ispartof Physics in medicine & biology, 2018-01, Vol.63 (2), p.025013-025013
issn 0031-9155
1361-6560
1361-6560
language eng
recordid cdi_crossref_primary_10_1088_1361_6560_aaa30c
source MEDLINE; IOP Publishing Journals; Institute of Physics (IOP) Journals - HEAL-Link
subjects Brachytherapy
Brachytherapy - methods
dual-energy CT
Humans
Male
Monte Carlo dose calculation
Monte Carlo Method
Phantoms, Imaging
Prostatic Neoplasms - diagnostic imaging
Prostatic Neoplasms - radiotherapy
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted - methods
tissue characterization
Tomography, X-Ray Computed - methods
title Dosimetric impact of dual-energy CT tissue segmentation for low-energy prostate brachytherapy: a Monte Carlo study
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-30T00%3A45%3A57IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Dosimetric%20impact%20of%20dual-energy%20CT%20tissue%20segmentation%20for%20low-energy%20prostate%20brachytherapy:%20a%20Monte%20Carlo%20study&rft.jtitle=Physics%20in%20medicine%20&%20biology&rft.au=Remy,%20Charlotte&rft.date=2018-01-11&rft.volume=63&rft.issue=2&rft.spage=025013&rft.epage=025013&rft.pages=025013-025013&rft.issn=0031-9155&rft.eissn=1361-6560&rft.coden=PHMBA7&rft_id=info:doi/10.1088/1361-6560/aaa30c&rft_dat=%3Cproquest_cross%3E1979171654%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=1979171654&rft_id=info:pmid/29260727&rfr_iscdi=true