Humidified single-scattering albedometer (H-CAPS-PMSSA): Design, data analysis, and validation

We report the development and validation of a new humidified aerosol single-scattering albedometer to quantify the effects of water uptake on submicrometer particle optical properties. The instrument simultaneously measures in situ aerosol light extinction (σ ep ) and scattering (σ sp ) using a cavi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Aerosol science and technology 2021-06, Vol.55 (7), p.749-768
Hauptverfasser: Carrico, Christian M., Capek, Tyler J., Gorkowski, Kyle J., Lam, Jared T., Gulick, Sabina, Karacaoglu, Jaimy, Lee, James E., Dungan, Charlotte, Aiken, Allison C., Onasch, Timothy B., Freedman, Andrew, Mazzoleni, Claudio, Dubey, Manvendra K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the development and validation of a new humidified aerosol single-scattering albedometer to quantify the effects of water uptake on submicrometer particle optical properties. The instrument simultaneously measures in situ aerosol light extinction (σ ep ) and scattering (σ sp ) using a cavity-attenuated phase shift-single scattering albedo particulate matter (PM) monitor (CAPS-PM SSA , Aerodyne Research, Inc., Billerica, MA, USA). It retrieves by difference aerosol light absorption (σ ap ) and directly quantifies aerosol single-scattering albedo (SSA), the aerosol "brightness." We custom built a relative humidity (RH) control system using a water vapor-permeable membrane humidifier and coupled it to the CAPS-PM SSA to enable humidified aerosol observations. Our humidified instrument (H-CAPS-PM SSA ) overcomes problems with noise caused by mirror purge-flow humidification, heating, and characterizing cell RH. Careful angular truncation corrections in scattering, particularly for larger particles, were combined with empirical observations. Results show that the optimal operational size to be D p < 400 nm. The H-CAPS-PM SSA was evaluated with several pure single-component aerosols including ammonium sulfate ((NH 4 ) 2 SO 4 ), absorbing nigrosin, and levoglucosan, an organic biomass smoke tracer. The measured σ ep , σ sp , and the derived optical hygroscopicity parameter (κ) for size-selected ammonium sulfate are in good agreement with literature values. For dry size-selected nigrosin in the 100 
ISSN:0278-6826
1521-7388