Development of a miniaturized duoplasmatron ion source
A miniature sized duoplasmatron ion source mountable on a standard 34 mm diameter copper gasket flange was designed, and its performance was tested. A hollow cathode exited a stable discharge in the anode through a 3 mm diameter orifice of the intermediate electrode with less than 10 W discharge pow...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2020-01, Vol.91 (1), p.013505-013505 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A miniature sized duoplasmatron ion source mountable on a standard 34 mm diameter copper gasket flange was designed, and its performance was tested. A hollow cathode exited a stable discharge in the anode through a 3 mm diameter orifice of the intermediate electrode with less than 10 W discharge power for Ar. The extracted ion beam current was larger at a smaller discharge current and was maximum at 5 mA below, where the discharge became unstable. A Faraday cup located 200 mm downstream of the extractor detected 1 µA beam current at 5 mA discharge current with 390 V discharge voltage and 2 kV extraction potential. The source required higher discharge voltage for operation of hydrogen gas. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.5128752 |