Measurement error of surface-mounted fiber Bragg grating temperature sensor
Fiber Bragg grating (FBG) sensors are extensively used to measure surface temperatures. However, the temperature gradient effect of a surface-mounted FBG sensor is often overlooked. A surface-type temperature standard setup was prepared in this study to investigate the measurement errors of FBG temp...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2014-06, Vol.85 (6), p.064905-064905 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Fiber Bragg grating (FBG) sensors are extensively used to measure surface temperatures. However, the temperature gradient effect of a surface-mounted FBG sensor is often overlooked. A surface-type temperature standard setup was prepared in this study to investigate the measurement errors of FBG temperature sensors. Experimental results show that the measurement error of a bare fiber sensor has an obvious linear relationship with surface temperature, with the largest error achieved at 8.1 °C. Sensors packaged with heat conduction grease generate smaller measurement errors than do bare FBG sensors and commercial thermal resistors. Thus, high-quality packaged methods and proper modes of fixation can effectively improve the accuracy of FBG sensors in measuring surface temperatures. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4885463 |