Measurement error of surface-mounted fiber Bragg grating temperature sensor

Fiber Bragg grating (FBG) sensors are extensively used to measure surface temperatures. However, the temperature gradient effect of a surface-mounted FBG sensor is often overlooked. A surface-type temperature standard setup was prepared in this study to investigate the measurement errors of FBG temp...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2014-06, Vol.85 (6), p.064905-064905
Hauptverfasser: Yi, Liu, Zude, Zhou, Erlong, Zhang, Jun, Zhang, Yuegang, Tan, Mingyao, Liu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Fiber Bragg grating (FBG) sensors are extensively used to measure surface temperatures. However, the temperature gradient effect of a surface-mounted FBG sensor is often overlooked. A surface-type temperature standard setup was prepared in this study to investigate the measurement errors of FBG temperature sensors. Experimental results show that the measurement error of a bare fiber sensor has an obvious linear relationship with surface temperature, with the largest error achieved at 8.1 °C. Sensors packaged with heat conduction grease generate smaller measurement errors than do bare FBG sensors and commercial thermal resistors. Thus, high-quality packaged methods and proper modes of fixation can effectively improve the accuracy of FBG sensors in measuring surface temperatures.