Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg
Laser-induced breakdown spectroscopy (LIBS) is a powerful analytical technique for elemental analysis, but matrix effects can significantly affect its accuracy. In this work, we propose the use of calibration-free LIBS (CF-LIBS) with the one-point calibration (OPC) method to reduce matrix effects in...
Gespeichert in:
Veröffentlicht in: | Journal of analytical atomic spectrometry 2023-05, Vol.38 (5), p.1155-1163 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 1163 |
---|---|
container_issue | 5 |
container_start_page | 1155 |
container_title | Journal of analytical atomic spectrometry |
container_volume | 38 |
creator | Leva Borduchi, Luís Carlos Menegatti, Carlos Renato Bastos Pereira Milori, Débora Marcondes Izário Filho, Hélcio José Villas-Boas, Paulino Ribeiro |
description | Laser-induced breakdown spectroscopy (LIBS) is a powerful analytical technique for elemental analysis, but matrix effects can significantly affect its accuracy. In this work, we propose the use of calibration-free LIBS (CF-LIBS) with the one-point calibration (OPC) method to reduce matrix effects in the quantification of mercury (Hg) for seven soil and eleven landfill leachate samples, with Hg concentration ranging from 50 to 1000 mg kg
−1
. To improve sensitivity, we employed a double-pulse LIBS system with parameters optimized for each of the two sample matrices. A single soil sample was used to calibrate the OPC factors, which were then applied to the CF-LIBS calculations for the other samples. Using CF-LIBS corrected with the OPC method, we achieved an average error of 4.3% in the quantification of Hg for the two sample sets. This result indicates that these methods have the potential not only to reduce matrix effects but also to compensate for variations in LIBS system optimization for either matrix (soils and leachates) in the quantification of Hg. Therefore, the combination of these methods opens new possibilities for LIBS analyses, where the system can be optimized for each element and sample matrix, and the quantification can be performed with CF-LIBS, as long as the requirements for local thermodynamic equilibrium are fulfilled. Further studies with other elements, matrices, and measurement conditions are needed to validate and expand these findings.
Combining calibration-free LIBS with the one-point calibration method can mitigate matrix effects and compensate for variations in system optimization, enabling accurate Hg quantification in soil and landfill leachate samples. |
doi_str_mv | 10.1039/d2ja00399f |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1039_D2JA00399F</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2811626720</sourcerecordid><originalsourceid>FETCH-LOGICAL-c281t-77972bc071a0e792d93bff2851a9f40a175fef5563246660821be26acaf0a4f3</originalsourceid><addsrcrecordid>eNpNkcFOwzAMhiMEEmNw4Y4UiRtSIUnbZOU2BoOhSRzYvXKzhGXqmi5JBXsJnpmMIuDkX_ZnW_6N0Dkl15Skxc2SrYFEUegDNKApz5I8z7JDNCCMi6TIhDhGJ96vCSFZzvIB-hy3bW0kBGMbbDW2jUpaa5qAJdSmcn1hPrt7xdo6vO2gCUb_a4AG6l1QHpsGe9i0dZTvJqzw0vhgGhnwBoIzH1iuwIEMyu3z0t9iiCu8wj50y13f8vR2io401F6d_cQhWkwfFpOnZP7yOJuM54lkIxoSIQrBKkkEBaJEwZZFWmnNRjmFQmcEqMi10nnOU5ZxzsmI0UoxDhI0gUynQ3TZj22d3XbKh3JtOxcv8WWcT3k0i5FIXfWUdNZ7p3TZOrMBtyspKfd2l_fsefxt9zTCFz3svPzl_t6RfgHboX5R</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2811626720</pqid></control><display><type>article</type><title>Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg</title><source>Alma/SFX Local Collection</source><source>Royal Society of Chemistry E-Journals</source><creator>Leva Borduchi, Luís Carlos ; Menegatti, Carlos Renato ; Bastos Pereira Milori, Débora Marcondes ; Izário Filho, Hélcio José ; Villas-Boas, Paulino Ribeiro</creator><creatorcontrib>Leva Borduchi, Luís Carlos ; Menegatti, Carlos Renato ; Bastos Pereira Milori, Débora Marcondes ; Izário Filho, Hélcio José ; Villas-Boas, Paulino Ribeiro</creatorcontrib><description>Laser-induced breakdown spectroscopy (LIBS) is a powerful analytical technique for elemental analysis, but matrix effects can significantly affect its accuracy. In this work, we propose the use of calibration-free LIBS (CF-LIBS) with the one-point calibration (OPC) method to reduce matrix effects in the quantification of mercury (Hg) for seven soil and eleven landfill leachate samples, with Hg concentration ranging from 50 to 1000 mg kg
−1
. To improve sensitivity, we employed a double-pulse LIBS system with parameters optimized for each of the two sample matrices. A single soil sample was used to calibrate the OPC factors, which were then applied to the CF-LIBS calculations for the other samples. Using CF-LIBS corrected with the OPC method, we achieved an average error of 4.3% in the quantification of Hg for the two sample sets. This result indicates that these methods have the potential not only to reduce matrix effects but also to compensate for variations in LIBS system optimization for either matrix (soils and leachates) in the quantification of Hg. Therefore, the combination of these methods opens new possibilities for LIBS analyses, where the system can be optimized for each element and sample matrix, and the quantification can be performed with CF-LIBS, as long as the requirements for local thermodynamic equilibrium are fulfilled. Further studies with other elements, matrices, and measurement conditions are needed to validate and expand these findings.
Combining calibration-free LIBS with the one-point calibration method can mitigate matrix effects and compensate for variations in system optimization, enabling accurate Hg quantification in soil and landfill leachate samples.</description><identifier>ISSN: 0267-9477</identifier><identifier>EISSN: 1364-5544</identifier><identifier>DOI: 10.1039/d2ja00399f</identifier><language>eng</language><publisher>London: Royal Society of Chemistry</publisher><subject>Analytical chemistry ; Calibration ; Error correction ; Laser induced breakdown spectroscopy ; Leachates ; Local thermodynamic equilibrium ; Mathematical analysis ; Optimization ; Parameter sensitivity</subject><ispartof>Journal of analytical atomic spectrometry, 2023-05, Vol.38 (5), p.1155-1163</ispartof><rights>Copyright Royal Society of Chemistry 2023</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c281t-77972bc071a0e792d93bff2851a9f40a175fef5563246660821be26acaf0a4f3</citedby><cites>FETCH-LOGICAL-c281t-77972bc071a0e792d93bff2851a9f40a175fef5563246660821be26acaf0a4f3</cites><orcidid>0000-0001-5225-0675 ; 0000-0002-8583-3954</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>Leva Borduchi, Luís Carlos</creatorcontrib><creatorcontrib>Menegatti, Carlos Renato</creatorcontrib><creatorcontrib>Bastos Pereira Milori, Débora Marcondes</creatorcontrib><creatorcontrib>Izário Filho, Hélcio José</creatorcontrib><creatorcontrib>Villas-Boas, Paulino Ribeiro</creatorcontrib><title>Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg</title><title>Journal of analytical atomic spectrometry</title><description>Laser-induced breakdown spectroscopy (LIBS) is a powerful analytical technique for elemental analysis, but matrix effects can significantly affect its accuracy. In this work, we propose the use of calibration-free LIBS (CF-LIBS) with the one-point calibration (OPC) method to reduce matrix effects in the quantification of mercury (Hg) for seven soil and eleven landfill leachate samples, with Hg concentration ranging from 50 to 1000 mg kg
−1
. To improve sensitivity, we employed a double-pulse LIBS system with parameters optimized for each of the two sample matrices. A single soil sample was used to calibrate the OPC factors, which were then applied to the CF-LIBS calculations for the other samples. Using CF-LIBS corrected with the OPC method, we achieved an average error of 4.3% in the quantification of Hg for the two sample sets. This result indicates that these methods have the potential not only to reduce matrix effects but also to compensate for variations in LIBS system optimization for either matrix (soils and leachates) in the quantification of Hg. Therefore, the combination of these methods opens new possibilities for LIBS analyses, where the system can be optimized for each element and sample matrix, and the quantification can be performed with CF-LIBS, as long as the requirements for local thermodynamic equilibrium are fulfilled. Further studies with other elements, matrices, and measurement conditions are needed to validate and expand these findings.
Combining calibration-free LIBS with the one-point calibration method can mitigate matrix effects and compensate for variations in system optimization, enabling accurate Hg quantification in soil and landfill leachate samples.</description><subject>Analytical chemistry</subject><subject>Calibration</subject><subject>Error correction</subject><subject>Laser induced breakdown spectroscopy</subject><subject>Leachates</subject><subject>Local thermodynamic equilibrium</subject><subject>Mathematical analysis</subject><subject>Optimization</subject><subject>Parameter sensitivity</subject><issn>0267-9477</issn><issn>1364-5544</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNpNkcFOwzAMhiMEEmNw4Y4UiRtSIUnbZOU2BoOhSRzYvXKzhGXqmi5JBXsJnpmMIuDkX_ZnW_6N0Dkl15Skxc2SrYFEUegDNKApz5I8z7JDNCCMi6TIhDhGJ96vCSFZzvIB-hy3bW0kBGMbbDW2jUpaa5qAJdSmcn1hPrt7xdo6vO2gCUb_a4AG6l1QHpsGe9i0dZTvJqzw0vhgGhnwBoIzH1iuwIEMyu3z0t9iiCu8wj50y13f8vR2io401F6d_cQhWkwfFpOnZP7yOJuM54lkIxoSIQrBKkkEBaJEwZZFWmnNRjmFQmcEqMi10nnOU5ZxzsmI0UoxDhI0gUynQ3TZj22d3XbKh3JtOxcv8WWcT3k0i5FIXfWUdNZ7p3TZOrMBtyspKfd2l_fsefxt9zTCFz3svPzl_t6RfgHboX5R</recordid><startdate>20230510</startdate><enddate>20230510</enddate><creator>Leva Borduchi, Luís Carlos</creator><creator>Menegatti, Carlos Renato</creator><creator>Bastos Pereira Milori, Débora Marcondes</creator><creator>Izário Filho, Hélcio José</creator><creator>Villas-Boas, Paulino Ribeiro</creator><general>Royal Society of Chemistry</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7SR</scope><scope>7U5</scope><scope>8BQ</scope><scope>8FD</scope><scope>JG9</scope><scope>L7M</scope><orcidid>https://orcid.org/0000-0001-5225-0675</orcidid><orcidid>https://orcid.org/0000-0002-8583-3954</orcidid></search><sort><creationdate>20230510</creationdate><title>Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg</title><author>Leva Borduchi, Luís Carlos ; Menegatti, Carlos Renato ; Bastos Pereira Milori, Débora Marcondes ; Izário Filho, Hélcio José ; Villas-Boas, Paulino Ribeiro</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c281t-77972bc071a0e792d93bff2851a9f40a175fef5563246660821be26acaf0a4f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Analytical chemistry</topic><topic>Calibration</topic><topic>Error correction</topic><topic>Laser induced breakdown spectroscopy</topic><topic>Leachates</topic><topic>Local thermodynamic equilibrium</topic><topic>Mathematical analysis</topic><topic>Optimization</topic><topic>Parameter sensitivity</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Leva Borduchi, Luís Carlos</creatorcontrib><creatorcontrib>Menegatti, Carlos Renato</creatorcontrib><creatorcontrib>Bastos Pereira Milori, Débora Marcondes</creatorcontrib><creatorcontrib>Izário Filho, Hélcio José</creatorcontrib><creatorcontrib>Villas-Boas, Paulino Ribeiro</creatorcontrib><collection>CrossRef</collection><collection>Engineered Materials Abstracts</collection><collection>Solid State and Superconductivity Abstracts</collection><collection>METADEX</collection><collection>Technology Research Database</collection><collection>Materials Research Database</collection><collection>Advanced Technologies Database with Aerospace</collection><jtitle>Journal of analytical atomic spectrometry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Leva Borduchi, Luís Carlos</au><au>Menegatti, Carlos Renato</au><au>Bastos Pereira Milori, Débora Marcondes</au><au>Izário Filho, Hélcio José</au><au>Villas-Boas, Paulino Ribeiro</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg</atitle><jtitle>Journal of analytical atomic spectrometry</jtitle><date>2023-05-10</date><risdate>2023</risdate><volume>38</volume><issue>5</issue><spage>1155</spage><epage>1163</epage><pages>1155-1163</pages><issn>0267-9477</issn><eissn>1364-5544</eissn><abstract>Laser-induced breakdown spectroscopy (LIBS) is a powerful analytical technique for elemental analysis, but matrix effects can significantly affect its accuracy. In this work, we propose the use of calibration-free LIBS (CF-LIBS) with the one-point calibration (OPC) method to reduce matrix effects in the quantification of mercury (Hg) for seven soil and eleven landfill leachate samples, with Hg concentration ranging from 50 to 1000 mg kg
−1
. To improve sensitivity, we employed a double-pulse LIBS system with parameters optimized for each of the two sample matrices. A single soil sample was used to calibrate the OPC factors, which were then applied to the CF-LIBS calculations for the other samples. Using CF-LIBS corrected with the OPC method, we achieved an average error of 4.3% in the quantification of Hg for the two sample sets. This result indicates that these methods have the potential not only to reduce matrix effects but also to compensate for variations in LIBS system optimization for either matrix (soils and leachates) in the quantification of Hg. Therefore, the combination of these methods opens new possibilities for LIBS analyses, where the system can be optimized for each element and sample matrix, and the quantification can be performed with CF-LIBS, as long as the requirements for local thermodynamic equilibrium are fulfilled. Further studies with other elements, matrices, and measurement conditions are needed to validate and expand these findings.
Combining calibration-free LIBS with the one-point calibration method can mitigate matrix effects and compensate for variations in system optimization, enabling accurate Hg quantification in soil and landfill leachate samples.</abstract><cop>London</cop><pub>Royal Society of Chemistry</pub><doi>10.1039/d2ja00399f</doi><tpages>9</tpages><orcidid>https://orcid.org/0000-0001-5225-0675</orcidid><orcidid>https://orcid.org/0000-0002-8583-3954</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0267-9477 |
ispartof | Journal of analytical atomic spectrometry, 2023-05, Vol.38 (5), p.1155-1163 |
issn | 0267-9477 1364-5544 |
language | eng |
recordid | cdi_crossref_primary_10_1039_D2JA00399F |
source | Alma/SFX Local Collection; Royal Society of Chemistry E-Journals |
subjects | Analytical chemistry Calibration Error correction Laser induced breakdown spectroscopy Leachates Local thermodynamic equilibrium Mathematical analysis Optimization Parameter sensitivity |
title | Application of one-point calibration LIBS for quantification of analytes in samples with distinct matrix characteristics: a case study with Hg |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-07T19%3A04%3A09IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Application%20of%20one-point%20calibration%20LIBS%20for%20quantification%20of%20analytes%20in%20samples%20with%20distinct%20matrix%20characteristics:%20a%20case%20study%20with%20Hg&rft.jtitle=Journal%20of%20analytical%20atomic%20spectrometry&rft.au=Leva%20Borduchi,%20Lu%C3%ADs%20Carlos&rft.date=2023-05-10&rft.volume=38&rft.issue=5&rft.spage=1155&rft.epage=1163&rft.pages=1155-1163&rft.issn=0267-9477&rft.eissn=1364-5544&rft_id=info:doi/10.1039/d2ja00399f&rft_dat=%3Cproquest_cross%3E2811626720%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2811626720&rft_id=info:pmid/&rfr_iscdi=true |