Pixel Value Modification Using RVG-4 Automatic Exposure Compensation for Instant High-Contrast Images

The RVG-4 permits automatic exposure compensation (AEC). The purpose of this investigation was to determine the effects of AEC on image contrast. Images were made either with or without a dental QA jaw phantom using a fixed image projection geometry. Exposures were 6.3 through 27.3 μC/kg using an X-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Oral Radiology 1996-06, Vol.12 (1), p.11-17
Hauptverfasser: Hayakawa, Yoshihiko, Farman, Allan G., Scarfe, William C., Kuroyanagi, Kinya
Format: Artikel
Sprache:eng ; jpn
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The RVG-4 permits automatic exposure compensation (AEC). The purpose of this investigation was to determine the effects of AEC on image contrast. Images were made either with or without a dental QA jaw phantom using a fixed image projection geometry. Exposures were 6.3 through 27.3 μC/kg using an X-ray generator operated at 70 kVp. Region of interest pixel value distributions were measured at tissue thicknesses in this phantom, and the average pixel values and signal to noise ratios (SNR) were calculated. The use of AEC without an object in place resulted in a disproportionate relationship between pixel value and exposure with a marked reduction in SNR. The use of AEC on under- and over -exposed images of the phantom simultaneously enhanced image contrast and reduced SNR. Thus, AEC provides a convenient and quick method for achieving high-contrast images with sub-optimal exposures, however, this could lead to inappropriate patient dosages if the function is used for over-exposed images. AEC reduces the SNR and produces disproportionate pixel values relative to exposure.
ISSN:0911-6028
1613-9674
DOI:10.1007/bf02351577