Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic

Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Marini, Marco, Capponi, Lorenzo, Battistoni, Ilaria, Tocci, Tommaso, Tribbiani, Giulio, Merante, Francesco, Francioni, Matteo, Matassini, Maria Vittoria, Pongetti, Giulia, Skhoza, Matilda, Angelini, Luca, Marsili, Roberto, Mazzanti, Marco, Perna, Gian Piero, Rossi, Gianluca, Gili, Alberto
Format: Artikel
Sprache:eng
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue
container_start_page
container_title
container_volume
creator Marini, Marco
Capponi, Lorenzo
Battistoni, Ilaria
Tocci, Tommaso
Tribbiani, Giulio
Merante, Francesco
Francioni, Matteo
Matassini, Maria Vittoria
Pongetti, Giulia
Skhoza, Matilda
Angelini, Luca
Marsili, Roberto
Mazzanti, Marco
Perna, Gian Piero
Rossi, Gianluca
Gili, Alberto
description Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.
doi_str_mv 10.48550/arxiv.2205.04167
format Article
fullrecord <record><control><sourceid>arxiv_GOX</sourceid><recordid>TN_cdi_arxiv_primary_2205_04167</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2205_04167</sourcerecordid><originalsourceid>FETCH-LOGICAL-a677-2f6aa151d97b35b35b087ae48de8c2a5afea47f6a11de57cf88e4c2d7ccd545f3</originalsourceid><addsrcrecordid>eNo1j81qwzAQhHXpoaR9gJ66L2BXsiVLOQb3LxCID6FXs5VWrSCWg-SU9u3rpA0MDAMzAx9jd4KX0ijFHzB9h6-yqrgquRSNvma2o-THNGC0BJgz5TxQnGD0MJALFveA0UEcY3HJbbfqIMRpHqKlDMdMDtwxhfgB0ydBu31bPxZiCYd5SUOwN-zK4z7T7b8v2O75ade-Fpvty7pdbQpstC4q3yAKJdxSv9fqJG40kjSOjK1QoSeUei4J4Uhp640haSunrXVKKl8v2P3f7ZmyP6QwYPrpT7T9mbb-BYmtUHE</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic</title><source>arXiv.org</source><creator>Marini, Marco ; Capponi, Lorenzo ; Battistoni, Ilaria ; Tocci, Tommaso ; Tribbiani, Giulio ; Merante, Francesco ; Francioni, Matteo ; Matassini, Maria Vittoria ; Pongetti, Giulia ; Skhoza, Matilda ; Angelini, Luca ; Marsili, Roberto ; Mazzanti, Marco ; Perna, Gian Piero ; Rossi, Gianluca ; Gili, Alberto</creator><creatorcontrib>Marini, Marco ; Capponi, Lorenzo ; Battistoni, Ilaria ; Tocci, Tommaso ; Tribbiani, Giulio ; Merante, Francesco ; Francioni, Matteo ; Matassini, Maria Vittoria ; Pongetti, Giulia ; Skhoza, Matilda ; Angelini, Luca ; Marsili, Roberto ; Mazzanti, Marco ; Perna, Gian Piero ; Rossi, Gianluca ; Gili, Alberto</creatorcontrib><description>Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.</description><identifier>DOI: 10.48550/arxiv.2205.04167</identifier><language>eng</language><creationdate>2022-05</creationdate><rights>http://arxiv.org/licenses/nonexclusive-distrib/1.0</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>228,230,776,881</link.rule.ids><linktorsrc>$$Uhttps://arxiv.org/abs/2205.04167$$EView_record_in_Cornell_University$$FView_record_in_$$GCornell_University$$Hfree_for_read</linktorsrc><backlink>$$Uhttps://doi.org/10.48550/arXiv.2205.04167$$DView paper in arXiv$$Hfree_for_read</backlink></links><search><creatorcontrib>Marini, Marco</creatorcontrib><creatorcontrib>Capponi, Lorenzo</creatorcontrib><creatorcontrib>Battistoni, Ilaria</creatorcontrib><creatorcontrib>Tocci, Tommaso</creatorcontrib><creatorcontrib>Tribbiani, Giulio</creatorcontrib><creatorcontrib>Merante, Francesco</creatorcontrib><creatorcontrib>Francioni, Matteo</creatorcontrib><creatorcontrib>Matassini, Maria Vittoria</creatorcontrib><creatorcontrib>Pongetti, Giulia</creatorcontrib><creatorcontrib>Skhoza, Matilda</creatorcontrib><creatorcontrib>Angelini, Luca</creatorcontrib><creatorcontrib>Marsili, Roberto</creatorcontrib><creatorcontrib>Mazzanti, Marco</creatorcontrib><creatorcontrib>Perna, Gian Piero</creatorcontrib><creatorcontrib>Rossi, Gianluca</creatorcontrib><creatorcontrib>Gili, Alberto</creatorcontrib><title>Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic</title><description>Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.</description><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><sourceid>GOX</sourceid><recordid>eNo1j81qwzAQhHXpoaR9gJ66L2BXsiVLOQb3LxCID6FXs5VWrSCWg-SU9u3rpA0MDAMzAx9jd4KX0ijFHzB9h6-yqrgquRSNvma2o-THNGC0BJgz5TxQnGD0MJALFveA0UEcY3HJbbfqIMRpHqKlDMdMDtwxhfgB0ydBu31bPxZiCYd5SUOwN-zK4z7T7b8v2O75ade-Fpvty7pdbQpstC4q3yAKJdxSv9fqJG40kjSOjK1QoSeUei4J4Uhp640haSunrXVKKl8v2P3f7ZmyP6QwYPrpT7T9mbb-BYmtUHE</recordid><startdate>20220509</startdate><enddate>20220509</enddate><creator>Marini, Marco</creator><creator>Capponi, Lorenzo</creator><creator>Battistoni, Ilaria</creator><creator>Tocci, Tommaso</creator><creator>Tribbiani, Giulio</creator><creator>Merante, Francesco</creator><creator>Francioni, Matteo</creator><creator>Matassini, Maria Vittoria</creator><creator>Pongetti, Giulia</creator><creator>Skhoza, Matilda</creator><creator>Angelini, Luca</creator><creator>Marsili, Roberto</creator><creator>Mazzanti, Marco</creator><creator>Perna, Gian Piero</creator><creator>Rossi, Gianluca</creator><creator>Gili, Alberto</creator><scope>GOX</scope></search><sort><creationdate>20220509</creationdate><title>Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic</title><author>Marini, Marco ; Capponi, Lorenzo ; Battistoni, Ilaria ; Tocci, Tommaso ; Tribbiani, Giulio ; Merante, Francesco ; Francioni, Matteo ; Matassini, Maria Vittoria ; Pongetti, Giulia ; Skhoza, Matilda ; Angelini, Luca ; Marsili, Roberto ; Mazzanti, Marco ; Perna, Gian Piero ; Rossi, Gianluca ; Gili, Alberto</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-a677-2f6aa151d97b35b35b087ae48de8c2a5afea47f6a11de57cf88e4c2d7ccd545f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><toplevel>online_resources</toplevel><creatorcontrib>Marini, Marco</creatorcontrib><creatorcontrib>Capponi, Lorenzo</creatorcontrib><creatorcontrib>Battistoni, Ilaria</creatorcontrib><creatorcontrib>Tocci, Tommaso</creatorcontrib><creatorcontrib>Tribbiani, Giulio</creatorcontrib><creatorcontrib>Merante, Francesco</creatorcontrib><creatorcontrib>Francioni, Matteo</creatorcontrib><creatorcontrib>Matassini, Maria Vittoria</creatorcontrib><creatorcontrib>Pongetti, Giulia</creatorcontrib><creatorcontrib>Skhoza, Matilda</creatorcontrib><creatorcontrib>Angelini, Luca</creatorcontrib><creatorcontrib>Marsili, Roberto</creatorcontrib><creatorcontrib>Mazzanti, Marco</creatorcontrib><creatorcontrib>Perna, Gian Piero</creatorcontrib><creatorcontrib>Rossi, Gianluca</creatorcontrib><creatorcontrib>Gili, Alberto</creatorcontrib><collection>arXiv.org</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext_linktorsrc</fulltext></delivery><addata><au>Marini, Marco</au><au>Capponi, Lorenzo</au><au>Battistoni, Ilaria</au><au>Tocci, Tommaso</au><au>Tribbiani, Giulio</au><au>Merante, Francesco</au><au>Francioni, Matteo</au><au>Matassini, Maria Vittoria</au><au>Pongetti, Giulia</au><au>Skhoza, Matilda</au><au>Angelini, Luca</au><au>Marsili, Roberto</au><au>Mazzanti, Marco</au><au>Perna, Gian Piero</au><au>Rossi, Gianluca</au><au>Gili, Alberto</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic</atitle><date>2022-05-09</date><risdate>2022</risdate><abstract>Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.</abstract><doi>10.48550/arxiv.2205.04167</doi><oa>free_for_read</oa></addata></record>
fulltext fulltext_linktorsrc
identifier DOI: 10.48550/arxiv.2205.04167
ispartof
issn
language eng
recordid cdi_arxiv_primary_2205_04167
source arXiv.org
title Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-05T18%3A10%3A40IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-arxiv_GOX&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20assessment%20of%20medical%20and%20non-medical%20CPAP%20interfaces%20used%20during%20the%20COVID-19%20pandemic&rft.au=Marini,%20Marco&rft.date=2022-05-09&rft_id=info:doi/10.48550/arxiv.2205.04167&rft_dat=%3Carxiv_GOX%3E2205_04167%3C/arxiv_GOX%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true