Studies of helium poisoning of a Hamamatsu R5900-00-M16 photomultiplier
We report results from studies of the helium poisoning of a 16-anode photomultiplier tube R5900-00-M16 manufactured by Hamamatsu Photonics. A tube was immersed in pure helium for a period of about four months and was periodically monitored using a digital oscilloscope. Our results are based on the a...
Gespeichert in:
Hauptverfasser: | , , , , , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report results from studies of the helium poisoning of a 16-anode
photomultiplier tube R5900-00-M16 manufactured by Hamamatsu Photonics. A tube
was immersed in pure helium for a period of about four months and was
periodically monitored using a digital oscilloscope. Our results are based on
the analysis of waveforms triggered by the dark noise pulses. Collected data
yield evidence of after-pulses due to helium contamination of the tube. The
probability of after-pulsing increased linearly with the exposure time to
helium but the phototube suffered only a small drop in gain, indicating
generally strong resilience to helium poisoning. |
---|---|
DOI: | 10.48550/arxiv.1908.08869 |