Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems

Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:arXiv.org 2018-07
Hauptverfasser: Segers, Tim, Stehouwer, Marco C, Filip M J J de Somer, de Mol, Bastian A, Versluis, Michel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue
container_start_page
container_title arXiv.org
container_volume
creator Segers, Tim
Stehouwer, Marco C
Filip M J J de Somer
de Mol, Bastian A
Versluis, Michel
description Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy of two bubble counters using monodisperse bubbles in a highly controlled setting at low GME concentrations. Monodisperse GME with a radius of 43 {\mu}m were produced in a microfluidic chip. Directly after their formation, they were injected one-by-one into the BCC200 and the EDAC sensors. GME size and count were optically verified using high-speed imaging. During best-case scenarios or low GME concentrations of GME with a size of 43 {\mu}m in radius in an in-vitro setup, the BCC200 overestimates GME size by a factor of 2 to 3 while the EDAC underestimates the average GME size by at least a factor of two. The BCC200 overestimates the GME concentration by approximately 20% while the EDAC overestimates the concentration by nearly one order of magnitude. Nevertheless, the calculated total GME volume is only over-predicted by a factor 2 since the EDAC underestimates the actual GME size. For the BCC200, the total GME volume was over-predicted by 25 times due to the over-estimation of GME size. The measured errors in the absolute sizing/counting of GME do not imply that all results obtained using the bubble counters are insignificant or invalid. A relative change in bubble size or bubble concentration can accurately be measured. However, care must be taken in the interpretation of the results and their absolute values. Moreover, the devices cannot be used interchangeably when reporting GME activity. Nevertheless, both devices can be used to study the relative air removal characteristics of CPB components or for the quantitative monitoring of GME production during CPB interventions.
doi_str_mv 10.48550/arxiv.1807.08946
format Article
fullrecord <record><control><sourceid>proquest_arxiv</sourceid><recordid>TN_cdi_arxiv_primary_1807_08946</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2092803389</sourcerecordid><originalsourceid>FETCH-LOGICAL-a529-641ca192b7f0e0e78f5c65391a156ae0c0abfc79a2ec9aa2d219d3971802f6fe3</originalsourceid><addsrcrecordid>eNotUMtqwzAQFIVCQ5oP6KmCnu3qYdnWsYS-IJBL7mYtS1TBtlzJdut-Qz-6StLT7uzMDsMgdEdJmpVCkEfw33ZOaUmKlJQyy6_QinFOkzJj7AZtQjgSQlheMCH4Cv3uh9EqaPGsvTVxG63rMfQNtn0y29E7rD7Agxoj_3NhncHjV7y7rtNeWWjbBcMMtoW61RiUm0L0xPVUn3CEfXwO2DiPFfjGumFqO9eDX3C9DBACDksYdRdu0bWBNujN_1yjw8vzYfuW7Pav79unXQKCySTPqAIqWV0YookuSiNULrikQEUOmigCtVGFBKaVBGANo7LhsoiVMJMbzdfo_mJ7rqoavO1ilupUWXWuLCoeLorBu89Jh7E6usn3MVPFiGQl4byU_A8_JnOD</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2092803389</pqid></control><display><type>article</type><title>Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems</title><source>arXiv.org</source><source>Free E- Journals</source><creator>Segers, Tim ; Stehouwer, Marco C ; Filip M J J de Somer ; de Mol, Bastian A ; Versluis, Michel</creator><creatorcontrib>Segers, Tim ; Stehouwer, Marco C ; Filip M J J de Somer ; de Mol, Bastian A ; Versluis, Michel</creatorcontrib><description>Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy of two bubble counters using monodisperse bubbles in a highly controlled setting at low GME concentrations. Monodisperse GME with a radius of 43 {\mu}m were produced in a microfluidic chip. Directly after their formation, they were injected one-by-one into the BCC200 and the EDAC sensors. GME size and count were optically verified using high-speed imaging. During best-case scenarios or low GME concentrations of GME with a size of 43 {\mu}m in radius in an in-vitro setup, the BCC200 overestimates GME size by a factor of 2 to 3 while the EDAC underestimates the average GME size by at least a factor of two. The BCC200 overestimates the GME concentration by approximately 20% while the EDAC overestimates the concentration by nearly one order of magnitude. Nevertheless, the calculated total GME volume is only over-predicted by a factor 2 since the EDAC underestimates the actual GME size. For the BCC200, the total GME volume was over-predicted by 25 times due to the over-estimation of GME size. The measured errors in the absolute sizing/counting of GME do not imply that all results obtained using the bubble counters are insignificant or invalid. A relative change in bubble size or bubble concentration can accurately be measured. However, care must be taken in the interpretation of the results and their absolute values. Moreover, the devices cannot be used interchangeably when reporting GME activity. Nevertheless, both devices can be used to study the relative air removal characteristics of CPB components or for the quantitative monitoring of GME production during CPB interventions.</description><identifier>EISSN: 2331-8422</identifier><identifier>DOI: 10.48550/arxiv.1807.08946</identifier><language>eng</language><publisher>Ithaca: Cornell University Library, arXiv.org</publisher><subject>Bubbles ; Counting ; Microfluidics ; Physics - Medical Physics ; Sizing</subject><ispartof>arXiv.org, 2018-07</ispartof><rights>2018. This work is published under http://arxiv.org/licenses/nonexclusive-distrib/1.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><rights>http://arxiv.org/licenses/nonexclusive-distrib/1.0</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>228,230,780,784,885,27925</link.rule.ids><backlink>$$Uhttps://doi.org/10.1177/0267659117722595$$DView published paper (Access to full text may be restricted)$$Hfree_for_read</backlink><backlink>$$Uhttps://doi.org/10.48550/arXiv.1807.08946$$DView paper in arXiv$$Hfree_for_read</backlink></links><search><creatorcontrib>Segers, Tim</creatorcontrib><creatorcontrib>Stehouwer, Marco C</creatorcontrib><creatorcontrib>Filip M J J de Somer</creatorcontrib><creatorcontrib>de Mol, Bastian A</creatorcontrib><creatorcontrib>Versluis, Michel</creatorcontrib><title>Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems</title><title>arXiv.org</title><description>Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy of two bubble counters using monodisperse bubbles in a highly controlled setting at low GME concentrations. Monodisperse GME with a radius of 43 {\mu}m were produced in a microfluidic chip. Directly after their formation, they were injected one-by-one into the BCC200 and the EDAC sensors. GME size and count were optically verified using high-speed imaging. During best-case scenarios or low GME concentrations of GME with a size of 43 {\mu}m in radius in an in-vitro setup, the BCC200 overestimates GME size by a factor of 2 to 3 while the EDAC underestimates the average GME size by at least a factor of two. The BCC200 overestimates the GME concentration by approximately 20% while the EDAC overestimates the concentration by nearly one order of magnitude. Nevertheless, the calculated total GME volume is only over-predicted by a factor 2 since the EDAC underestimates the actual GME size. For the BCC200, the total GME volume was over-predicted by 25 times due to the over-estimation of GME size. The measured errors in the absolute sizing/counting of GME do not imply that all results obtained using the bubble counters are insignificant or invalid. A relative change in bubble size or bubble concentration can accurately be measured. However, care must be taken in the interpretation of the results and their absolute values. Moreover, the devices cannot be used interchangeably when reporting GME activity. Nevertheless, both devices can be used to study the relative air removal characteristics of CPB components or for the quantitative monitoring of GME production during CPB interventions.</description><subject>Bubbles</subject><subject>Counting</subject><subject>Microfluidics</subject><subject>Physics - Medical Physics</subject><subject>Sizing</subject><issn>2331-8422</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2018</creationdate><recordtype>article</recordtype><sourceid>ABUWG</sourceid><sourceid>AFKRA</sourceid><sourceid>AZQEC</sourceid><sourceid>BENPR</sourceid><sourceid>CCPQU</sourceid><sourceid>DWQXO</sourceid><sourceid>GOX</sourceid><recordid>eNotUMtqwzAQFIVCQ5oP6KmCnu3qYdnWsYS-IJBL7mYtS1TBtlzJdut-Qz-6StLT7uzMDsMgdEdJmpVCkEfw33ZOaUmKlJQyy6_QinFOkzJj7AZtQjgSQlheMCH4Cv3uh9EqaPGsvTVxG63rMfQNtn0y29E7rD7Agxoj_3NhncHjV7y7rtNeWWjbBcMMtoW61RiUm0L0xPVUn3CEfXwO2DiPFfjGumFqO9eDX3C9DBACDksYdRdu0bWBNujN_1yjw8vzYfuW7Pav79unXQKCySTPqAIqWV0YookuSiNULrikQEUOmigCtVGFBKaVBGANo7LhsoiVMJMbzdfo_mJ7rqoavO1ilupUWXWuLCoeLorBu89Jh7E6usn3MVPFiGQl4byU_A8_JnOD</recordid><startdate>20180724</startdate><enddate>20180724</enddate><creator>Segers, Tim</creator><creator>Stehouwer, Marco C</creator><creator>Filip M J J de Somer</creator><creator>de Mol, Bastian A</creator><creator>Versluis, Michel</creator><general>Cornell University Library, arXiv.org</general><scope>8FE</scope><scope>8FG</scope><scope>ABJCF</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>AZQEC</scope><scope>BENPR</scope><scope>BGLVJ</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>HCIFZ</scope><scope>L6V</scope><scope>M7S</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>PTHSS</scope><scope>GOX</scope></search><sort><creationdate>20180724</creationdate><title>Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems</title><author>Segers, Tim ; Stehouwer, Marco C ; Filip M J J de Somer ; de Mol, Bastian A ; Versluis, Michel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-a529-641ca192b7f0e0e78f5c65391a156ae0c0abfc79a2ec9aa2d219d3971802f6fe3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2018</creationdate><topic>Bubbles</topic><topic>Counting</topic><topic>Microfluidics</topic><topic>Physics - Medical Physics</topic><topic>Sizing</topic><toplevel>online_resources</toplevel><creatorcontrib>Segers, Tim</creatorcontrib><creatorcontrib>Stehouwer, Marco C</creatorcontrib><creatorcontrib>Filip M J J de Somer</creatorcontrib><creatorcontrib>de Mol, Bastian A</creatorcontrib><creatorcontrib>Versluis, Michel</creatorcontrib><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>Materials Science &amp; Engineering Collection</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central Essentials</collection><collection>ProQuest Central</collection><collection>Technology Collection</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>SciTech Premium Collection</collection><collection>ProQuest Engineering Collection</collection><collection>Engineering Database</collection><collection>Publicly Available Content Database</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>Engineering Collection</collection><collection>arXiv.org</collection><jtitle>arXiv.org</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Segers, Tim</au><au>Stehouwer, Marco C</au><au>Filip M J J de Somer</au><au>de Mol, Bastian A</au><au>Versluis, Michel</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems</atitle><jtitle>arXiv.org</jtitle><date>2018-07-24</date><risdate>2018</risdate><eissn>2331-8422</eissn><abstract>Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy of two bubble counters using monodisperse bubbles in a highly controlled setting at low GME concentrations. Monodisperse GME with a radius of 43 {\mu}m were produced in a microfluidic chip. Directly after their formation, they were injected one-by-one into the BCC200 and the EDAC sensors. GME size and count were optically verified using high-speed imaging. During best-case scenarios or low GME concentrations of GME with a size of 43 {\mu}m in radius in an in-vitro setup, the BCC200 overestimates GME size by a factor of 2 to 3 while the EDAC underestimates the average GME size by at least a factor of two. The BCC200 overestimates the GME concentration by approximately 20% while the EDAC overestimates the concentration by nearly one order of magnitude. Nevertheless, the calculated total GME volume is only over-predicted by a factor 2 since the EDAC underestimates the actual GME size. For the BCC200, the total GME volume was over-predicted by 25 times due to the over-estimation of GME size. The measured errors in the absolute sizing/counting of GME do not imply that all results obtained using the bubble counters are insignificant or invalid. A relative change in bubble size or bubble concentration can accurately be measured. However, care must be taken in the interpretation of the results and their absolute values. Moreover, the devices cannot be used interchangeably when reporting GME activity. Nevertheless, both devices can be used to study the relative air removal characteristics of CPB components or for the quantitative monitoring of GME production during CPB interventions.</abstract><cop>Ithaca</cop><pub>Cornell University Library, arXiv.org</pub><doi>10.48550/arxiv.1807.08946</doi><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier EISSN: 2331-8422
ispartof arXiv.org, 2018-07
issn 2331-8422
language eng
recordid cdi_arxiv_primary_1807_08946
source arXiv.org; Free E- Journals
subjects Bubbles
Counting
Microfluidics
Physics - Medical Physics
Sizing
title Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T07%3A33%3A50IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_arxiv&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Optical%20verification%20and%20in-vitro%20characterization%20of%20two%20commercially%20available%20acoustic%20bubble%20counters%20for%20cardiopulmonary%20bypass%20systems&rft.jtitle=arXiv.org&rft.au=Segers,%20Tim&rft.date=2018-07-24&rft.eissn=2331-8422&rft_id=info:doi/10.48550/arxiv.1807.08946&rft_dat=%3Cproquest_arxiv%3E2092803389%3C/proquest_arxiv%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2092803389&rft_id=info:pmid/&rfr_iscdi=true