The design, construction, and preliminary testing of a second generation proton CT scanner is presented. All current treatment planning systems at proton therapy centers use X-ray CT as the primary imaging modality for treatment planning to calculate doses to tumor and healthy tissues. One of the li...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , , , , , , , |
---|---|
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The design, construction, and preliminary testing of a second generation
proton CT scanner is presented. All current treatment planning systems at
proton therapy centers use X-ray CT as the primary imaging modality for
treatment planning to calculate doses to tumor and healthy tissues. One of the
limitations of X-ray CT is in the conversion of X-ray attenuation coefficients
to relative (proton) stopping powers, or RSP. This results in more proton range
uncertainty, larger target volumes and therefore, more dose to healthy tissues.
To help improve this, we present a novel scanner capable of high dose rates, up
to 2~MHz, and large area coverage, 20~x~24~cm$^2$, for imaging an adult head
phantom and reconstructing more accurate RSP values. |
---|---|
DOI: | 10.48550/arxiv.1409.0049 |